01711544172
bmibd.net@gmail.com
৳6,700.00

Hematology analyzer LX304HMA 5 part differentiation fully automatic benchtop unit that can carry out 60 tests per hour. Designed with touch screen display, Integral titanium incubation system with sample mode is autoloader mode with auto mixing, random direction tube barcode scanning function. Working principle has Laser Scatter Technology. It features an automatically calculates CV% and saves data and stores up to 100,000 results including histograms and scattergram. It is used widely in patient and research. All the function are high efficient and easy and fast to operate.



Share with:
Throughput 60Tests/hour
Assay Items 5 parts, 28 parameters, 2 histograms, 2 scattergrams
Test mode CBC+5DIFF, CBC
Measuring Principle Laser Scatter Technology
Sample Modes Autoloader: 50 tubes, auto mixing, random direction tube barcode scanning technology
Close tube: STAT priority, support both whole blood and capillary blood samples
Testing Items Regular Parameters:WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, PDW, MPV, PCT, P-LCR, NEU#, NEU%, LYM#, LYM%, MON#, EOS#, EOS%, BAS#, BAS%, Research Parameters:ALY#, ALY%, IG#, IG%
Linearity Range
Item Linearity range Carry Over CV
WBC 1.0~99.9x10^9/L ≤1.0% ≤2.0%
RBC 0.3~7.0x10^12/L ≤1.0% ≤1.5%
HGB 20~240g/L ≤1.0% ≤1.5%
PLT 20~999x10^9/L ≤1.0% ≤4.0%
Storage 100,000 results storage including histogram and scattergram
LIS Interface Support HL7 protocol
QC and Calibration Multiple QC rules,including L-J,X-B etc. Auto and manual calibration funtion
Incubation Integral titanium incubation system
Multi-warning Reagent inventory check; Abnormal original data warning; Multi warning flags
Power supply AC220V±10% 50/60Hz; 110V±10% 60Hz; 600VA
Operation environment 15~30℃
Package Size (W x D x H) 820 x 770 x 840 mm
Gross weight 110 kgs